- Contact Person : Ms. Chen Iry
- Company Name : Wuxi Jinyibo Instrument Technology Co., Ltd.
- Tel : 86-510-83217963
- Fax : 86-510-83217963
- Address : Jiangsu,Wuxi,No.35 Jingshen Rd, Qianqiao Development Zone
- Country/Region : China
- Zip : 214151
TY-9610 Type Photoelectric Direct Reading Spectrometer
Main technical parameters:
1. Test Matrix (Multi - matrix): Fe,Al,Cu,Ni,Co,Mg,Ti,Zn,Pb,Sn and Ag, etc. Sample analysis of various metals as well as their alloy.
2. Analysis Passageway (Multi- passageway): 45 passageways.
3. Analysis Waveband (Wide extent): 160nm~650nm
4. Testing Time (Short time): on the basis of different sample type, usually less than 30s.
5. Optics Room (Refined and constant temperature): 38°C ± 0.2°C
6. Power Source: protective grounding one-phase power source. Voltage: 220VAC Frequency: 50Hz Service power: 1.8KVA
7. Environment Condition: working temperature: 10-30°C
storage temperature: 0-45°C humidity: ≤85%
Main allocation:
1. Optical system
Structure: Palmer-Runge
Grating:
Radius of curvature: 750mm
Scribing density: 2400 lines/mm
Scribing area: 30×50mm2
Blaze wavelength (1st order): 300nm
Line dispersion: 0.55nm/mm
Incident slit width: 20μm
Emergent slit width: 50μm/75μm
Photomultiplier: 28mm dia., 10 level side window tube, fused quartz
Allow passageway set: 45 passageways
Local constant temperature of spectrometer: 38°C ± 0.2°C
2. Vacuum measurement and control
Automatic control by the vacuum valve, to guarantee the vacuum stability of the light room, improve the stability of the optical system, and decrease the pollution of oil vapor to light room greatly.
Stored program control of the vacuum system completely, guarantees the vacuum degree and reduces the performance time of vacuum pump meanwhile, extend the service life of the vacuum pump.
The linkage of high-pressure open and vacuum system of PMT(Photo Multiplier Tube) can prevent glow discharging.
Exciting light is guided into vacuum room by lens, eliminate the loss of light path and keep the long-time stability of the data.
3. Sample excitation table device
Sample exciting table design is open type, it can analyze samples with different sizes and shapes.
The analytical gap of sample table: 4mm, excitation room is small.
The height of compression bar can be adjusted freely, easy to exchange samples.
Optimized argon gas circuit design, low consumption, sample excitation is more sufficient, discharging is more concentrated.
4. Excitation light source
Input power: 1.0KVA
Charging capacitance: precombustion 6μF, exposure 1.5μF
Peak current: precombustion 120A, exposure 30A
Peak voltage in main circuit: 300VDC (automatic regulation of peak voltage)
Shot-firing circuit: pulse amplitude:+15KV
auxiliary gap:adopt tunnel diode
discharge frequency: 150Hz--600Hz be adjustable
5. Measurement and control system
Measurement and control system: exchange data with the upper computer directly.
Measurement mode: subsection integral.
Capacity of integral capacitor: 0.047μF
Reproducibility: RSD≤0.2%
Voltage: -1000V
Stability: superior to 0.5% within 8h.
HV regulation: change from -550V to -1000V, regulating in 8(0-7) grades.
HV setup: altered by users via computer software.
6. Analysis software
Working under WINDOWS XP operating system, the main function is as follows:
1)single key shortcut operation analysis and other task operation.
2)automatic analysis program.
3)manual data input and output.
4)flexible analysis results display and print.
5)quality inspect and classify.
6)recalculation of the concentration result.
7)instrument standardization.
8)type standardization.
9)software and instrument allocation tools and application.
10)result storage.
11)flexible sample definition.
12)user right crypto guard.
13)data processing tool.
7. External device
1)brand commercial microcomputer
2)stylus printer or laser printer
3)hardware dog
The new appearance shape design now.
The appearance shape design before.
TY-9610 Type Photoelectric Direct Reading Spectrometer